Dosimetric effects of supine immobilization devices on the skin in intensity-modulated radiation therapy for breast cancer: a retrospective study

Abstract

Background

The dose perturbation effect of immobilization devices is often overlooked in intensity-modulated radiation therapy (IMRT) for breast cancer (BC). This retrospective study assessed the dosimetric effects of supine immobilization devices on the skin using a commercial treatment planning system.

Methods

Forty women with BC were divided into four groups according to the type of primary surgery: groups A and B included patients with left and right BC, respectively, who received 50 Gy radiotherapy in 25 fractions after radical mastectomy, while groups C and D included patients with left and right BC, respectively, who received breast-conservation surgery (BCS) and 40.05 Gy in 15 fractions as well as a tumor bed simultaneous integrated boost to 45 Gy. A 0.2-cm thick skin contour and two sets of body contours were outlined for each patient. Dose calculations were conducted for the two sets of contours using the same plan. The dose differences were assessed by comparing the dose-volume histogram parameter results and by plan subtraction.

Results

The supine immobilization devices for BC resulted in significantly increased skin doses, which may ultimately lead to skin toxicity. The mean dose increased by approximately 0.5 and 0.45 Gy in groups A and B after radical mastectomy and by 2.7 and 3.25 Gy in groups C and D after BCS; in groups A–D, the percentages of total normal skin volume receiving equal to or greater than 5 Gy (V5) increased by 0.54, 1.15, 2.67, and 1.94%, respectively, while the V10 increased by 1.27, 1.83, 1.36, and 2.88%; the V20 by 0.85, 1.87, 2.76, and 4.86%; the V30 by 1.3, 1.24, 10.58, and 11.91%; and the V40 by 1.29, 0.65, 10, and 10.51%. The dose encompassing the planning target volume and other organs at risk, showed little distinction between IMRT plans without and with consideration of immobilization devices.

Conclusions

The supine immobilization devices significantly increased the dose to the skin, especially for patients with BCS. Thus, immobilization devices should be included in the external contour to account for dose attenuation and skin dose increment.

Trial registration

This study does not report on interventions in human health care.

Peer Review reports

Background

Breast immobilization devices are commonly used in radiation oncology to provide breast cancer (BC) patients support and improve positional reproducibility during their fractionated radiotherapy [1,2,3]. In actual clinical practice, the beam attenuation and build-up perturbation effect caused by immobilization devices are often overlooked because the carbon fiber materials widely used in these devices are believed to be radio translucent for mega-voltage photons [4]. However, the density of carbon fiber is not equivalent to air; thus, attenuation and scattering can occur when the radiation beams pass cross these immobilization systems [5, 6]. Previous studies have reported that immobilization devices used in radiotherapy reduced the tumor dose, increased the skin dose (bolus effect), and altered the dose distributions [7,8,9]. De Puysseleyr and colleagues reported that irradiating through carbon fiber immobilization devices for prone breast radiotherapy resulted in considerable beam attenuation (range: 5.33 to 7.57%) and degradation of skin sparing [7]. For Chinese BC patients, due to their small and compact breast glands, supine positioning remains the most common approach and has multiple advantages, including methodological simplicity, comfort and accuracy, reproducible positioning, and reduced mean dose to the heart [10, 11].

Compared to conventional wedge-based breast radiotherapy, intensity-modulated radiation therapy (IMRT) can deliver highly conformal and homogeneous dose distributions to targets and, thus, significantly decrease clinical toxicities such as dermatitis and edema [12,13,14]. These advantages have substantially increased the adoption of IMRT during breast radiotherapy [15, 16]. However, the increased beams and monitor units (MUs) have an increased propensity to deliver radiation beams through the immobilization devices, resulting in radiation immobilization device attenuation, ultimately compromising the target coverage and organ-at-risk (OAR) protection [17]. However, no study has yet assessed the dosimetric effects of supine breast immobilization devices on the delivered doses to the target volume and OARs for breast IMRT. Thus, this study quantified the dosimetric effect of supine immobilization devices by comparing the dose distributions calculated with and without breast immobilization devices and investigated the potential skin sparing for BC patients achievable with 6 MV photon beams in IMRT plans.

Methods

Patient data and setup

This study enrolled 40 women with BC who received adjuvant radiotherapy in our institution. The participants were divided into four groups according to the lesion location, type of primary site surgery, and irradiation field. Patients with left or right BC receiving radical mastectomy and chest plus infra/supraclavicular lymph node irradiation were assigned into groups A and B respectively. To ensure the same irradiation fields, the patients in these two groups had four or more involved lymph nodes and required infra/supraclavicular lymph node irradiation. In the same way, patients with left or right BC receiving breast-conservation surgery (BCS) and breast without lymph region irradiation were divided into groups C or D, respectively. No metastatic lymph nodes were detected in these two group patients and only the breast was irradiated. The patients’ ages ranged from 32 to 65 years, with a median age of 47 years.

Simulation

All patients were simulated in the head-first supine position using a carbon fiber breast bracket (Klarity Inc., Guangzhou, China) for body immobilization. The supporting board was inclined at 7°, 12°, 17°, or 23° to assure that the sternum was horizontal. The patients’ heads were positioned straight on a circle sponge head support, with the chin slightly upwards, avoiding skin folds at the lower neck. Both arms were raised over their heads using a pair of arm supports to adequately expose the breast, as well as a knee support to prevent the body from sliding down. A thermoplastic film (electron density 0.3–0.7, thickness 2.4 mm) (Klarity Medical Products, Newark OH) was custom-molded over the chest and attached to the bracket with a plastic batten (electron density 1–1.1). Computed tomography (CT) imaging with a 3-mm slice thickness was performed using a large-aperture CT simulation scanner (Brilliance, Philips Medical System, Amsterdam, Netherlands) (Fig.1). The scan range was from the first cervical vertebra to the diaphragm. The simulation CT images were transferred to the treatment planning system (TPS, Monaco V5.11, Elekta AB, Stockholm, Sweden) for target and OAR delineation and treatment planning.

Fig. 1
figure 1

Display of the immobilization devices in the axial (a) and sagittal (b) views. The orange portion is the couch, the purple portion is the breast-board, the green portion is the chest fixation mask of thermoplastic, the skin contour is displayed in yellow, and the PTV is red

Regions of interest (ROIs)

The ROIs were delineated on CT images with the CT data set as soft tissue (window 600, level 40) by experienced oncologists according to the recommendation of the International Commission on Radiation Units and Measurements (ICRU Reports 83). For patients who underwent BCS, the clinical target volumes (CTVs) included all mammary glandular tissue and a CTV boost (the tumor bed including the clips and seroma plus a 5-mm margin in all directions without exceeding the CTV of the breast). The corresponding planning target volumes (PTVs) were generated by uniformly expanding 5 mm from the CTVs. Given the smaller mammary glands of Eastern women, the ventral border was placed 2 mm below the skin surface [18]. For patients with modified radical mastectomy, the CTV included the chest wall and infra/supraclavicular lymph nodes, with the ventral border next to the skin surface. The OARs, including the heart, left anterior descending artery (LAD), left ventricle (LV), contralateral breast, ipsilateral lung, and contralateral lung, were accounted for in left-side BC, while the liver was especially accounted for in right-side BC. The thyroid, larynx, esophagus, and spinal cord were also considered in patients receiving lymph node irradiation. To assess the surface dose variance from immobilization devices in TPS, the skin contour in this study was especially delineated in the treatment region with 2 mm thickness below the skin surface for each patient [Fig.1] [19].

For commercial TPS, an external structure, which should contain the materials involved in the calculation, must be defined to calculate the dose distributions. In this study, two sets of external body contours were created for each patient: one set included only the patient’s body without immobilization devices, while the other set included the patient’s external body contours and the whole breast immobilization device.

Treatment planning and dose calculation

The prescription doses to the PTV boost and PTV breast were 45 Gy and 40.05 Gy, respectively, with a total of 15 fractions in the case of BCS. For mastectomy, the prescription was 50 Gy in 25 fractions. All patients were planned on a Monaco TPS at 6 MV using the dynamic inverse IMRT technique. Multi-beam IMRT employs three groups of similarly opposed lateral fields spaced through a 290–150° sector for left-side tumors and a 200–60° sector for right-side tumors around the target volume, which includes the breast/chest wall and regional nodes, as indicated in Fig.1. A 0° field was added when the periclavicular node region was included. A 0.5-cm bolus was added to the surface skin in the treatment region for patients with radical mastectomy to compensate for the build-up effect of X-rays; in contrast, a bolus was avoided for patients with breast-conserving surgery to improve the cosmetic outcome. The optimization was performed using the Monaco’s build-in Monte Carlo (MC) algorithm combined with Dynamic Multi-Leaf Collimator (DMLC) technology. The maximum and minimum doses were planned in accordance with the ICRU 83 recommendations, with dose constraints following the QUANTEC directive [19,20,21].

Two IMRT plans were generated for each patient, with plans not including the immobilization devices in the calculations defined as Plan- and dose distributions recalculated with the external body contour containing the immobilization device under the same irradiation constraints defined as Plan+.

Statistical analysis

Dose-volume histograms (DVHs) are the popular method to evaluate the dose coverage of PTVs and OARs. For the PTVs in the present study, the parameters were the mean dose (Dmean), the homogeneity index (HI), and the conformity index (CI). The HI and CI were respectively calculated as follows [22, 23].

$$ HI=\frac{D_{5\%}}{D_{95\%}} $$
(1)
$$ CI=\frac{TV1}{TV}\ast \frac{TV1}{VR1} $$
(2)

In formula (1), D5% and D95% were the doses received by 5 and 95% of the ROI volume, respectively. A HI value closer to 1 indicates a better uniformity of the dose distribution in the target volume. In formula (2), TV1 is the target volume that receives the prescription dose while TV is the target volume. VR1 is the total volume within the prescription isodose curve. The CI value ranges between 0 and 1, with higher values indicating better dose conformity. Regarding OARs, the average dose Dmean, as well as the maximum dose (Dmax) and the dose-volume were calculated.

For each patient, the dosimetric effects due to the immobilization devices were calculated by plan subtraction in the TPS. \( \overline{D} \) represented the average of parameter differences between Plan+ and Plan-, as shown in formula 3, while \( \overline{D} \) % represents the average of the relative differences between Plan+ and Plan-.

$$ \overline{D}={\sum}_1^{10}\left[\left( plan+\right)-\left( plan-\right)\right]/10 $$
(3)
$$ \overline{D}\%={\sum}_1^{10}\left\{\left[\left( plann+\right)-\left( plann-\right)\right]\ast 100/\left( plann+\right)\right\}/10 $$
(4)

IBM SPSS Statistics for Windows, version 22.0 (IBM Corp., Armonk, NY, USA) was used to analyze all data. Wilcoxon matched-paired signed-rank tests were used to evaluate the significance of the observed differences between Plan+ and Plan-. The differences were considered statistically significant when p < 0.05.

Results

The comparisons of dosimetric differences between Plan+ and Plan- for BC patients receiving radical mastectomy are presented in Tables 1 and 2. The parameters (Coverage Index, Dmean, D2%, and CI) of the PTV showed little difference, except for the HI of left-side cancer (\( \overline{D} \) = − 0.006) and D98% of right-side cancer (\( \overline{D} \) = − 0.38 Gy) with statistically insignificant differences. Due to the bolus effect of the breast immobilization devices, the mean dose and relative volumes of skin receiving 5, 10, 20, 30, and 40 Gy were significantly increased for Plan+ (\( \overline{D} \) and \( \overline{D} \) % of 0.50 Gy and 1.25, 0.54 and 0.56%, 1.27 and 1.37%, 0.85 and 1.00%, 1.30 and 1.67%, and 1.29 and 1.99% for left BC and 0.45 Gy and 1.11, 1.15 and 1.20%, 1.83 and 2.01%, 1.87 and 2.27%, 1.24 and 1.56%, and 0.65 and 0.96% for right BC, respectively, all p < 0.05, Fig. 2a). However, there was no statistically significant difference in other OARs, except for the V5 of the larynx for right BC (\( \overline{D} \) = 6.94 Gy, p = 0.014).

Table 1 Dosimetric parameters of PTV and OARs for 10 cases of left breast cancer after radical mastectomy
Table 2 Dosimetric parameters of PTV and OARs for 10 cases of right breast cancer after radical mastectomy
Fig. 2
figure 2

The \( \overline{D} \) % of skin dosimetries for breast cancer after radical mastectomy (a) and BCS (b). The error bars reflect the standard error of the mean (r/ √n). The lines are drawn only to guide the eye

For patients with breast-conserving surgery, the dosimetric effects of immobilization devices were also calculated between Plan+ and Plan-. As shown in Tables 3 and 4 and Fig. 2b, the plans calculation with breast immobilization devices showed higher mean skin doses (\( \overline{D} \) % = 9.07 ± 1.60% for left BC and 10.21 ± 2.95% for right BC) and higher volumes of skin receiving dose (5–40 Gy) radiation (\( \overline{D} \) % 2.91, 1.65, 3.57, 15.85, and 51.86% for left BC and 2.05, 3.17, 5.84, 16.49 and 51.63% for right BC, respectively, Fig. 2b), which again displayed the bolus effect of immobilization devices. For left-side BC patients, the mean dose and relative irradiation volume of 5-30Gy of the left lung decreased, with little clinically significance, in Plan+ (\( \overline{D} \)-0.21 Gy for Dmean, − 1.18% for V5, − 0.98% for V10, − 0.47% for V20, and − 0.28% for V30, respectively). Regarding PTV, the coverage index, HI, and CI were also altered, with statistical but not clinical significance. The other OARs far from the PTV, such as the contra-lung, heart, LV, LAD, and liver, showed non-significant differences between the two plans.

Table 3 Dosimetric parameters of PTV and OARs for 10 cases of left breast cancer after breast-conserving surgery
Table 4 Dosimetric parameters of PTV and OARs for 10 cases of right breast cancer after breast-conserving surgery

Dose difference distribution map (plan+ − Plan-)

The dose difference distributions were calculated as Plan+ subtracted from Plan-. As shown in Fig. 3a, the blue to red gradient represented different absolute dose values ranging from − 5 to 5 Gy. The build-up effect and radiation scattering caused by the immobilization devices dramatically altered the dose distributions. The skin dose was observably increased in the irradiated region when the immobilization devices were included in the calculations. In other words, the skin dose was underestimated by approximately 6 Gy if the immobilization devices were not included in the external contour. The doses in other regions including Lung-L and PTV were also decreased, a finding similar to the DVH and data comparison results, as show in Fig.3b.

Fig. 3
figure 3

a Dose difference distributions of the cross-sectional plane for a typical patient with left-side breast cancer after BCS. The dose difference was calculated by subtracting Plan- (calculated without immobilization devices) from Plan+ (calculated with the whole immobilization devices included in the external body structure). b DVH results of Plan- and Plan+ for a typical patient with left-side breast cancer after BCS. The solid and dotted lines represent the results of Plan- and Plan+, respectively

Discussion

Patient immobilization devices have become an important tool to guarantee the accurate delivery of highly conformal dose distributions [3]. As the materials used in immobilization devices are not completely X-ray transmissible and can cause attenuation of the delivered dose, the dosimetric effects of immobilization devices should be included in dose calculations [17, 24]. Beam attenuation from the couch, additional inserts, and immobilization devices can cause a misrepresentation of the actual dose delivered to the PTV, with a deviation of more than the recommended 3–5% accuracy range reported by Olch [3]. Chen reported the attenuation of head and neck immobilization devices, which reduced the dose coverage rate from 1.51 to 9.92% and the average dose from 0.93 to 1.92% of planning target volumes in nasopharyngeal carcinoma [24]. Olson assessed the dose variance from immobilization devices in volumetric-modulated arc therapy (VMAT) head and neck treatment planning and found that the plan calculated without immobilization devices was problematic, showing compromised V95, D100, and PTV coverage [17]. However, in our study, we observed no clinically important effect of supine breast immobilization devices on the dosimetric parameters of PTV and PGTV, with a deviation of less than 3%. The potential reason for this difference may lie in the fact that not all radiation beams passed through the couch in our study. Puysseleyr et al. measured the dosimetric impact of a prone breast immobilization device and found that beam attenuation accounted for 7.6% (6 MV X-ray) of beams passing through the couch top-base plate combination and almost 5% for beams traversing the couch-top [7]. Thus, a beam attenuation of less than 3% occurred when the beam passed through the base plate only, similar to our findings.

In addition, the bolus effect cannot be avoided. Beams, especially posterior oblique orientations beams, passing through the immobilization devices involved in treatment can result in increased unexpected skin doses to ultimately affect the dosimetric effects [1, 8, 24, 25]. Kelly et al. utilized radiochromic film and MOSFET detectors to quantify the effect of an immobilization cast on skin dose in breast radiotherapy, and observed an increase in skin dose up to 45.7 and 62.30%, which is similar to our results of patients with breast-conserving surgery (\( \overline{D} \) % = 51.86 and 51.63% for left and right BC) [1]. Lee and colleagues measured the inguinal region skin dose using thermoluminescent dosimeters (TLDs) in prostate cancer patients. They found that the TLD-measured dose was two-fold higher than the calculated dose that did not contour the vacuum cushion and couch and was similar to the calculated dose with both devices contoured [8]. Chen et al. also reported that, due to the bolus effect of head and neck immobilization devices, the dorsal neck skin dose was significantly increased by approximately 8 Gy (53%) in multi-field IMRT for nasopharyngeal cancer [24]. Similarly, Ali et al. find that thermoplastic mask used to patients for head and neck, pelvis and thoracic treatment can significantly increase skin dose by up to a factor of 4 more than that without the mask using 6 MV beams [25]. Radiation-induced skin toxicity (RIST) is a predominant adverse effect and deserves consideration as severe skin toxicity can lead to treatment cession and cosmetic changes in BC patients. In the present study, we also observed a bolus effect of immobilization devices, as the skin mean dose and volume receiving 5–40 Gy were significantly increased in Plan+. This effect was more obvious in patients after BCS, mainly because the breast was spatially closer to the immobilization device compared to the chest. Moreover, the V30 appeared to be the most sensitive parameter except for in patients with right BC receiving radical mastectomy. Pastore et al. reported that breast skin receiving doses ≥30 Gy was the most predictive parameter of acute RIST [26], while Tsair-Fwu showed that skin receiving a dose > 35 Gy (V35) was the most significant dosimetric predictor associated with radiation dermatitis grade 2+ toxicity. The higher V30 in our study may translate into increased RIST. There is currently no standard of practice to include immobilization devices within body contours; the results of this study showed that the actual skin dose was underestimated when treatment beams passed through the couch top and immobilization devices, which, in turn, induced more and severer dermatitis. Our results indicate that immobilization devices should be included in dose calculations in BC treatment planning and that the skin of the breast region should be delineated as an OAR and that a dose-volume constraint for skin should be defined whenever possible.

Despite the positive results of this study, it has some limitations. A larger patient population, as well as different TPS or calculation algorithms, dosimetry techniques, and dose measurements are required in future studies.

Conclusions

This study calculated and evaluated the dosimetric effects on the skin of supine immobilization devices for BC in IMRT plans. The data showed a significantly increased skin dose, especially in patients with BCS, with both the V30 and V40 of the skin increasing sharply by more than 10%. These findings should remind radiation practitioners to pay attention to the skin dose caused by the immobilization devices and to seek solutions to ameliorate these negative effects. It is possible to include the immobilization devices within the external body contour and to account for the skin dose increment in the TPS calculation in BC treatment planning.

Availability of data and materials

The datasets used and/or analysed during the current study are available from the corresponding author on reasonable request.

Abbreviations

BC:

breast cancer

BCS:

breast-conservation surgery

CI:

conformity index

CSTRO:

China Society for Radiation Oncology

CT:

computed tomography

CTV:

clinical target volume

Dmean:

mean dose

DMLC:

Dynamic Multi-Leaf Collimator

DVH:

dose-volume histogram

HI:

homogeneity index

ICRU:

International Commission on Radiation Units and Measurements Reports

IMRT:

intensity-modulated radiation therapy

LAD:

left anterior descending artery

LV:

left ventricle

MC:

Monte Carlo

MUs:

monitor units

OAR:

organ at risk

PTV:

planning target volume

RIST:

radiation-induced skin toxicity

ROI:

region of interest

TLD:

thermoluminescent dosimeter

TPS:

treatment planning system

VMAT:

volumetric-modulated arc therapy

Vx:

volume of skin receiving equal to or greater than X Gy

References

  1. Kelly A, Hardcastle N, Metcalfe P, Cutajar D, Quinn A, Foo K, et al. Surface dosimetry for breast radiotherapy in the presence of immobilization cast material. Phys Med Biol. 2011;56(4):1001–13. https://0-doi-org.brum.beds.ac.uk/10.1088/0031-9155/56/4/008.

  2. Xiang Q, Jie W, Zhu K, Wang Q, Cheng J. Which technique of positioning and immobilization is better for breast cancer patients in postmastectomy IMRT, single-pole or double-pole immobilization? J Appl Clin Med Physics. 2019;20(1):168–74. https://0-doi-org.brum.beds.ac.uk/10.1002/acm2.12506.

    Article  Google Scholar 

  3. Olch AJ, Gerig L, Li H, Mihaylov I, Morgan A. Dosimetric effects caused by couch tops and immobilization devices: report of AAPM task group 176. Med Phys. 2014;41(6):061501. https://0-doi-org.brum.beds.ac.uk/10.1118/1.4876299.

    Article  PubMed  Google Scholar 

  4. Meara SJ, Langmack KA. An investigation into the use of carbon fibre for megavoltage radiotherapy applications. Phys Med Biol. 1998;43(5):1359–66. https://0-doi-org.brum.beds.ac.uk/10.1088/0031-9155/43/5/025.

    Article  CAS  PubMed  Google Scholar 

  5. Sheykhoo A, Abdollahi S, Hadizadeh Yazdi MH, Ghorbani M, Mohammadi M. Effects of Siemens TT-D carbon fiber table top on beam attenuation, and build up region of 6 MV photon beam. Reports Pract Oncol Radiother. 2017;22(1):19–28. https://0-doi-org.brum.beds.ac.uk/10.1016/j.rpor.2016.09.001.

    Article  Google Scholar 

  6. Vieira SC, Kaatee RS, Dirkx ML, Heijmen BJ. Two-dimensional measurement of photon beam attenuation by the treatment couch and immobilization devices using an electronic portal imaging device. Med Phys. 2003;30(11):2981–7. https://0-doi-org.brum.beds.ac.uk/10.1118/1.1620491.

    Article  PubMed  Google Scholar 

  7. De Puysseleyr A, De Neve W, De Wagter C. A patient immobilization device for prone breast radiotherapy: Dosimetric effects and inclusion in the treatment planning system. Phys Med. 2016;32(6):758–66. https://0-doi-org.brum.beds.ac.uk/10.1016/j.ejmp.2016.04.013.

    Article  PubMed  Google Scholar 

  8. Lee KW, Wu JK, Jeng SC, Hsueh Liu YW, Cheng JC. Skin dose impact from vacuum immobilization device and carbon fiber couch in intensity modulated radiation therapy for prostate cancer. Med Dosim. 2009;34(3):228–32. https://0-doi-org.brum.beds.ac.uk/10.1016/j.meddos.2008.10.001.

    Article  PubMed  Google Scholar 

  9. Munjal RK, Negi PS, Babu AG, Sinha SN, Anand AK, Kataria T. Impact of 6MV photon beam attenuation by carbon fiber couch and immobilization devices in IMRT planning and dose delivery. J Med Physics. 2006;31(2):67–71. https://0-doi-org.brum.beds.ac.uk/10.4103/0971-6203.26690.

    Article  CAS  Google Scholar 

  10. Mulliez T, Gulyban A, Vercauteren T, van Greveling A, Speleers B, De Neve W, et al. Setup accuracy for prone and supine whole breast irradiation. Strahlentherapie und Onkologie. 2016;192(4):254–9. https://0-doi-org.brum.beds.ac.uk/10.1007/s00066-016-0943-6.

    Article  PubMed  Google Scholar 

  11. Yu T, Xu M, Sun T, Shao Q, Zhang Y, Liu X, et al. External-beam partial breast irradiation in a supine versus prone position after breast-conserving surgery for Chinese breast cancer patients. Sci Rep. 2018;8(1):15354. https://0-doi-org.brum.beds.ac.uk/10.1038/s41598-018-33741-z.

  12. Harsolia A, Kestin L, Grills I, Wallace M, Jolly S, Jones C, et al. Intensity-modulated radiotherapy results in significant decrease in clinical toxicities compared with conventional wedge-based breast radiotherapy. Int J Radiat Oncol Biol Phys. 2007;68(5):1375–80. https://0-doi-org.brum.beds.ac.uk/10.1016/j.ijrobp.2007.02.044.

  13. Kestin LL, Sharpe MB, Frazier RC, Vicini FA, Yan D, Matter RC, et al. Intensity modulation to improve dose uniformity with tangential breast radiotherapy: initial clinical experience. Int J Radiat Oncol Biol Phys. 2000;48(5):1559–68. https://0-doi-org.brum.beds.ac.uk/10.1016/S0360-3016(00)01396-1.

  14. Vicini FA, Sharpe M, Kestin L, Martinez A, Mitchell CK, Wallace MF, et al. Optimizing breast cancer treatment efficacy with intensity-modulated radiotherapy. Int J Radiat Oncol Biol Phys. 2002;54(5):1336–44. https://0-doi-org.brum.beds.ac.uk/10.1016/S0360-3016(02)03746-X.

  15. Roberts KB, Soulos PR, Herrin J, Yu JB, Long JB, Dostaler E, et al. The adoption of new adjuvant radiation therapy modalities among Medicare beneficiaries with breast cancer: clinical correlates and cost implications. Int J Radiat Oncol Biol Phys. 2013;85(5):1186–92. https://0-doi-org.brum.beds.ac.uk/10.1016/j.ijrobp.2012.10.009.

  16. Wang EH, Mougalian SS, Soulos PR, Smith BD, Haffty BG, Gross CP, et al. Adoption of intensity modulated radiation therapy for early-stage breast cancer from 2004 through 2011. Int J Radiat Oncol Biol Phys. 2015;91(2):303–11. https://0-doi-org.brum.beds.ac.uk/10.1016/j.ijrobp.2014.09.011.

  17. Olson A, Phillips K, Eng T, Lenards N, Hunzeker A, Lewis D, et al. Assessing dose variance from immobilization devices in VMAT head and neck treatment planning: a retrospective case study analysis. Med Dosim. 2018;43(1):39–45. https://0-doi-org.brum.beds.ac.uk/10.1016/j.meddos.2017.08.001.

  18. Ma J, Li J, Xie J, Chen J, Zhu C, Cai G, et al. Post mastectomy linac IMRT irradiation of chest wall and regional nodes: dosimetry data and acute toxicities. Radiation Oncol (London, England). 2013;8:81.

    Article  Google Scholar 

  19. Lancellotta V, Iacco M, Perrucci E, Falcinelli L, Zucchetti C, de Bari B, et al. Comparing four radiotherapy techniques for treating the chest wall plus levels III-IV draining nodes after breast reconstruction. Br J Radiol. 2018;91(1086):20160874. https://0-doi-org.brum.beds.ac.uk/10.1259/bjr.20160874.

  20. Palumbo I, Mariucci C, Falcinelli L, Perrucci E, Lancellotta V, Podlesko AM, et al. Hypofractionated whole breast radiotherapy with or without hypofractionated boost in early stage breast cancer patients: a mono-institutional analysis of skin and subcutaneous toxicity. Breast Cancer (Tokyo, Japan). 2019;26(3):290–304.

    Article  Google Scholar 

  21. Lancellotta V, Chierchini S, Perrucci E, Saldi S, Falcinelli L, Iacco M, et al. Skin toxicity after chest wall/breast plus level III-IV lymph nodes treatment with helical tomotherapy. Cancer Investig. 2018;36(9–10):504–11. https://0-doi-org.brum.beds.ac.uk/10.1080/07357907.2018.1545854.

  22. Oliver M, Chen J, Wong E, Van Dyk J, Perera F. A treatment planning study comparing whole breast radiation therapy against conformal, IMRT and tomotherapy for accelerated partial breast irradiation. Radiother Oncol. 2007;82(3):317–23. https://0-doi-org.brum.beds.ac.uk/10.1016/j.radonc.2006.11.021.

    Article  PubMed  Google Scholar 

  23. Nakamura JL, Verhey LJ, Smith V, Petti PL, Lamborn KR, Larson DA, et al. Dose conformity of gamma knife radiosurgery and risk factors for complications. Int J Radiat Oncol Biol Phys. 2001;51(5):1313–9. https://0-doi-org.brum.beds.ac.uk/10.1016/S0360-3016(01)01757-6.

  24. Chen L, Peng YL, Gu SY, Shen H, Zhang DD, Sun WZ, et al. Dosimetric effects of head and neck immobilization devices on multi-field intensity modulated radiation therapy for nasopharyngeal carcinoma. J Cancer. 2018;9(14):2443–50. https://0-doi-org.brum.beds.ac.uk/10.7150/jca.24887.

  25. Ali I, Matthiesen C, Algan O, Thompson S, Bogardus C, Herman T, et al. Quantitative evaluation of increase in surface dose by immobilization thermoplastic masks and superficial dosimetry using Gafchromic EBT film and Monte Carlo calculations. J X-ray Sci Technol. 2010;18(3):319–26. https://0-doi-org.brum.beds.ac.uk/10.3233/XST-2010-0263.

  26. Pastore F, Conson M, D'Avino V, Palma G, Liuzzi R, Solla R, et al. Dose-surface analysis for prediction of severe acute radio-induced skin toxicity in breast cancer patients. Acta Oncol. 2016;55(4):466–73. https://0-doi-org.brum.beds.ac.uk/10.3109/0284186X.2015.1110253.

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

This work was financial supported by Science and Technology Planning Project of Quanzhou Science and Technology Bureau grant number [2018N001S] and Fujian Provincial Health and Family planning Commission Research Talent training project grant number [2018-1-60]. The funding bodies played no role in the design of the study and collection, analysis, and interpretation of data and in writing the manuscript.

Author information

Authors and Affiliations

Authors

Contributions

RL drafted the manuscript, conducted IMRT plans and collected the data. YW conceived and designed this study, contoured the reigns of interest, and revised the manuscript. GY conducted IMRT plans and performed the analysis. YH selected the patient collective and helped to conducted IMRT plans. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Yanhong Wang.

Ethics declarations

Ethics approval and consent to participate

As the data are anonymous and no intervention was happened in the treatment of patient, only verbal informed consent was obtained from participants, this procedure was approved by the Second Affiliated Hospital of Fujian Medical University Ethic Committee [2020(371)].

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no conflict of interest.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Lv, R., Yang, G., Huang, Y. et al. Dosimetric effects of supine immobilization devices on the skin in intensity-modulated radiation therapy for breast cancer: a retrospective study. BMC Cancer 21, 384 (2021). https://0-doi-org.brum.beds.ac.uk/10.1186/s12885-021-08119-6

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://0-doi-org.brum.beds.ac.uk/10.1186/s12885-021-08119-6

Keywords